| Title | Eye lens dose distribution in neonatal head CT: An analysis of the scanning conditions and CTDIvol limitations |
| Journal | Formal name:Radiation Protection Dosimetry Abbreviation:Radiat Prot Dosimetry |
| Domestic / Foregin | Foregin |
| Publisher | Oxford University Press |
| Volume, Issue, Page | ncag046 |
| Author and coauthor | Tatsuhiro Gotanda, Takuya Akagawa, Tomoyuki Hasuo, Shinnosuke Nishihara, Kohsei Matsuura, Yasuyuki Kawaji, Hidetoshi Yatake, Shinya Imai, Rumi Gotanda, Toshizo Katsuda |
| Authorship | Lead author,Corresponding author |
| Publication date | 2026/05/18 |
| Summary | Despite neonatal radiosensitivity, accurate lens dose assessments in computed tomography (CT) are lacking. We assessed the neonatal head CT eye lens dose under various scanning conditions, including vertical table displacement and organ-based exposure modulation (OEM), using the GAFCHROMIC LD-V1 (LD-V1), which can measure two-dimensional dose distributions, and we evaluated the limitations of conventional dose indices. The lens dose decreased with increasing distance from the isocentre for the Canon 320. Whereas, for the Canon 80 and Siemens, the dose peaked at +25 mm and then decreased. For the OEM, the dose reduction effect decreased at 30°. The CT dose index volume remained consistent. Spatially resolved dosimetry using the LD-V1 reveals beam overlap and positioning-related dose anomalies that cannot be captured by point dosimeters or conventional indices. The LD-V1 can effectively evaluate superficial dose distributions and guide dose optimization in neonatal imaging. |