| Title | Energy dependence of the GAFCHROMIC LD-V1 in the diagnostic radiographic modalities |
| Journal | Formal name:Journal of Applied Clinical Medical Physics ISSN code:15269914 |
| Domestic / Foregin | Foregin |
| Publisher | John Wiley & Sons, Inc |
| Volume, Issue, Page | 26(7),pp.e70117 |
| Author and coauthor | Tatsuhiro Gotanda, Tomoyuki Hasuo, Shinnosuke Nishihara, Kohsei Matsuura, Yasuyuki Kawaji, Hidetoshi Yatake, Shinya Imai, Takuya Akagawa, Nobuyoshi Tanki, Toshizo Katsuda |
| Authorship | Lead author,Corresponding author |
| Publication date | 2025/05/09 |
| Summary | The GAFCHROMIC LD-V1 radiochromic film is widely used in dosimetry because it can provide high-resolution two-dimensional dose distributions without processing. This study aimed to evaluate the response characteristics at different effective energies, from the low-energy range of mammography to the high-energy range of computed tomography. Net pixel value (NPV)-absorbed dose calibration curves for the GAFCHROMIC LD-V1 were generated using x-rays with effective energies of 18, 30, 50, and 80 keV to reflect those used in different diagnostic radiographic modalities. The film response was analyzed using calibration curves at each energy level. The coefficients of determination for the calibration curves at 18, 30, 50, and 80 keV were 0.9992, 0.9997, 0.9999, and 0.9976, respectively. The pixel value change at 30 keV was the largest and most sensitive, while the smallest change in pixel value and lowest sensitivity were noted at 18 keV. Because the energy dependence of the GAFCHROMIC LD-V1 is significant below 18 keV and above 80 keV, it is necessary to establish an appropriate NPV-absorbed dose calibration curve for energies below 18 keV and consider the possibility of underestimating the dose at energies above 80 keV. |
| DOI | https://doi.org/10.1002/acm2.70117 |