ゴタンダ タツヒロ   Tatsuhiro Gotanda
  五反田 龍宏
   所属   川崎医療福祉大学  医療技術学部 診療放射線技術学科
   職種   准教授
論文種別 原著
言語種別 英語
査読の有無 査読あり
表題 Three-Dimensional Dosimetry of Mammography Using a Block Cube Breast Phantom and Radiochromic Film
掲載誌名 正式名:Radiation Protection Dosimetry
略  称:Radiat Prot Dosimetry
ISSNコード:01448420
掲載区分国外
出版社 OXFORD ACADEMIC
巻・号・頁 196(3-4),pp.226-233
著者・共著者 Tatsuhiro Gotanda, Toshizo Katsuda, Hidetoshi Yatake, Yasuyuki Kawaji, Rumi Gotanda, Shinya Imai, Takuya Akagawa, Masashi Sasaki
担当区分 筆頭著者
発行年月 2021/10/20
概要 This study developed a phantom with a shape similar to that of the breast and use GAFCHROMIC films that can be placed inside the phantom to measure the detailed breast dose distribution in mammography. GAFCHROMIC EBT3 was placed on the block cube breast phantom (BCBP) and irradiated with a mammography device to measure the absorbed dose distribution inside the phantom in the horizontal and depth directions. The dose distribution in the horizontal plane was the highest in the centre on the chest wall side, and it decreased in a fan shape. Along the depth of the phantom, the doses absorbed across the entire cross-section were 16.15 mGy at the surface and 7.51, 3.25, and 1.68 mGy at depths of 10, 20, and 30 mm, respectively. Compared to the mean glandular dose, the proposed method can measure breast dose distributions in greater detail and is applicable to various breast shapes.
DOI https://doi.org/10.1093/rpd/ncab153