ツジ シュウヘイ   Shuhei Tsuji
  辻 修平
   所属   川崎医科大学  医学部 一般教養 自然科学
   職種   准教授
論文種別 原著
言語種別 英語
査読の有無 査読あり
表題 Monte Carlo dosimetric characterization of the IRAsource high dose rate Iridium-192 brachytherapy source: Comparison with the mHDR-v2r model
掲載誌名 正式名:PLOS One
掲載区分国外
巻・号・頁 21(2)
著者・共著者 Shuhei Tsuji
担当区分 筆頭著者,責任著者
発行年月 2026/02/06
概要 The aim of this study was to evaluate the dosimetric characteristics of an IRAsource
brachytherapy source using Monte Carlo simulations incorporating source electron
effects. The dose rate constant, radial dose function, and two-dimensional (2D)
anisotropy function were calculated and compared with those of published data and
results of the mHDR-v2r source model. The dose rate constant obtained for the IRAsource
was (1.1142 ± 0.0030) cGyh−1U−1, which was consistent with previously
reported values, within the range of experimental uncertainty. The radial dose function
exhibited a pattern similar to that of the mHDR-v2r source, with dose distributions
converging at distances of approximately 10 cm, where the effect of capsule
thickness became negligible. The 2D anisotropy function did not fully align with other
experimental datasets for the IRAsource. Even when comparing the IRAsource with
mHDR-v2r sources, the 2D anithotropy function showed a difference of 3% to 6%
around 0 degrees at distances of 0.25 cm to 5.0 cm from the source. The results
reflected reasonable trends that were consistent with differences in source capsule
geometry of the IRAsource and the mHDR-v2r. These findings show that the dosimetric
data presented here is reliable. This study provides essential baseline data for
accurate dose assessment of the IRAsource in brachytherapy and underscores the
importance of further experimental validation to refine dose characterization.
researchmap用URL https://doi.org/10.1371/journal.pone.0339841