ツジ シュウヘイ
Shuhei Tsuji
辻 修平 所属 川崎医科大学 医学部 一般教養 自然科学 職種 准教授 |
|
論文種別 | 原著 |
言語種別 | 英語 |
査読の有無 | 査読あり |
表題 | High dose rate 192Ir brachytherapy source model Monte Carlo dosimetry: mHDR-v2 and mHDR-v2r |
掲載誌名 | 正式名:PLOS ONE |
掲載区分 | 国外 |
巻・号・頁 | 19(2) |
総ページ数 | 17 |
著者・共著者 | Shuhei Tsuji, Naomasa Narihiro, Masataka Oita, Yoshihito Namito, Hideo Hirayama |
担当区分 | 筆頭著者,責任著者 |
発行年月 | 2024/02/09 |
概要 | After 2010, the source model of the microSelectron HDR Afterloader System was slightly modified from the previous model. Granero et al. named the modified source model “mHDR-v2r (revised model mHDR-v2)” and the previous model “mHDR-v2”. They concluded that the dosimetric differences arising from the dimensional changes between the mHDR-v2 and mHDR-v2r designs were negligible at almost all locations (within 0.5% for r >= 0.25 cm), the two-dimensional anisotropy function difference between the two sources is found 2.1% at r = 1.0 cm when compared with the results of the other experimental group. To confirm this difference, we performed a full Monte Carlo simulation without energy-fluence approximation. This is useful near the radiation source where charged-particle equilibrium does not hold. The two-dimensional anisotropy function of the TG-43U1 dataset showed a few percent difference between the mHDR-v2r and mHDR-v2 sources. There was no agreement in the immediate vicinity of the source (0.10 cm and 0.25 cm), when compared to Granero et al. in mHDR-v2r sources. The differences in these two-dimensional anisotropy functions were identified. |
researchmap用URL | https://doi.org/10.1371/journal.pone.0298550 |